MAGNETOM Trio, A Tim System 3T
|
|
|
|
|
|
|
The following information will provide you with more technical details about MAGNETOM Trio, A Tim System with I-class.
|
|
|
|
|
- Up to 102 seamlessly integrated coil elements with up to 32 RF channels
- 50 cm FoV. Whole-body imaging functionality up to 181 cm (5’11’’)
- iPAT2. Unmatched PAT up to 16
|
|
|
|
|
|
|
|
- Short bore 3T magnet
- Homogeneity: 0.25 ppm VRMS for a 40 cm (16 inch) DSV (diameter spherical volume)
1 ppm VRMS for a 50 cm
(20 inch) DSV (diameter spherical volume)
- Unmatched magnet reliability
|
|
|
|
|
|
|
|
- Hardware and Software measures for acoustic noise reduction up to 20 dB (A) as compared to conventional 3T systems. This is a reduction of 90 % in sound pressure.
- Gradient field strength up to 45 mT/m (72 mT/m effective)
- Slew rate up to 200 T/m/s (346 T/m/s effective)
- Large Field of View up to 50 cm / 20 inches with excellent linearity
- Ultrafast, highly compact, water-cooled gradient amplifier in solid-state technology for best min. TR 1.5 ms
and min. TE 0.6 ms (matrix 2562)
|
|
|
|
|
|
syngo, the highly intuitive Siemens cross-modality user interface, is the basis for all MAGNETOM Trio with I-class applications. Enabling you to expand your capabilities in every clinical field - from women's health to functional MRI.
|
|
|
|
|
- 33 m2 floor space only
- No computer room required for a total of just two electronic cabinets (water-cooled)
|
|
|
|
|
|
|
|
|
|
|
|